OneTouch diabetes control solution

diabetes control solution

Control Solution should be used at least once a week, whenever you suspect the meter or test strips may not be functioning properly, and every time a new vial of test strips is opened. Use of control solution is also recommended for teaching or learning your LifeScan System.
 

OneTouch® Verio® Control Solution

OneTouch® Verio® Control Solution is for use only with OneTouch® Verio® Test Strips.

Please read the packing insert and your  OneTouch®  Verio®IQ or OneTouch®  Verio® Owner’s Booklet before using OneTouch® Verio® Control Solution (available separately).

When to carry out a Control Solution test:

  • Whenever you open a new vial of OneTouch® Verio® Test Strips
  • If you suspect that the meter or test strips are not working properly
  • If you have had repeated unexpected blood glucose results
  • If you drop or damage the meter

Each vial contains enough solution for 75 tests.

A control solution test result that is within the expected range printed on the control solution vial will reassure you that the meter and test strips are working together and you are carrying out the test correctly.

OneTouch® Ultra® Control Solution

OneTouch® Ultra® Control Solution is for use only with OneTouch® Ultra® FastDraw™ Design Test Strips.

  • Read package insert before using this product.
  • The vial contains sufficient volume for 75 tests.
  • Refer to your OneTouch® Ultra® Meter Owner's Booklet for detailed control solution test instructions.

OneTouch® Vita® Control Solution

OneTouch® Vita® is for use only with OneTouch® Vita® Test Strips.

Please read the packing insert and your OneTouch® Vita® Use's Handbook before using OneTouch® Vita® Control Solution.

Carry out a control solution test:

  • To practice the test process, instead of using blood.
  • Once a week.
  • If you suspect that the meter or test strips are not working properly.
  • If you have had repeated unexpected blood glucose results.
  • If you drop or damage the meter.

 

Each vial contains enough solution for a minimum of 75 tests.

A control solution test that is within the expected range printed on the test strip vial will tell you that the meter and test strips are working together and you are doing the test properly.

OneTouch® Control Solution

OneTouch® Control Solution is for use only with OneTouch® Test Strips.

  • Read package insert before using this product.
  • The vial contains sufficient volume for 75 tests.
  • Refer to your LifeScan Meter Owner's Booklet for detailed control solution test instructions.

Control Solution Storage and Handling:

  • Discard 3 months after opening
  • Write the discard date on vial.
  • Do not use if expiration date has passed.
  • Store at temperatures under 30°C (86°F).
  • Do not refrigerate.

Precautions:

For in vitro diagnostics use only.
Shake vial well before use.